Same-day versus consecutive-day precision error of dual-energy X-ray absorptiometry for interpreting body composition change in resistance trained athletes
Journal article   Open access   Peer reviewed

Same-day versus consecutive-day precision error of dual-energy X-ray absorptiometry for interpreting body composition change in resistance trained athletes

Adam J Zemski, Karen Hind, Shelley E Keating, Elizabeth M Broad, Damian J Marsh and Gary J Slater
Journal of Clinical Densitometry, Vol.22(1), pp.104-114
2019
PMID: 30454952
pdf
PDF - Author Accepted Version (Open Access)587.42 kBDownloadView
Accepted VersionPDF - Author Accepted Version (Open Access)CC BY-NC-ND V4.0 Open Access
url
https://doi.org/10.1016/j.jocd.2018.10.005View
Published Version

Abstract

least significant change LSC DXA lean mass fat mass
Introduction: The application of dual-energy X-ray absorptiometry (DXA) in sport science settings is gaining popularity due to its ability to assess body composition. The International Society for Clinical Densitometry (ISCD) recommends application of the least significant change (LSC) to interpret meaningful and true change. This is calculated from same-day consecutive scans, thus accounting for technical error. However, this approach doesn't capture biological variation which is pertinent when interpreting longitudinal measurements, and could be captured from consecutive-day scans. The aims of this study were to investigate the impact short-term biological variation has on LSC measures, and establish if there is a difference in precision based on gender in a resistance trained population. Methodology: Twenty-one resistance trained athletes (age 30.6±8.2 years; stature 174.2±7.2cm; mass 74.3±11.6kg) with at least 12 months consistent resistance training experience, underwent two consecutive DXA scans on one day of testing, and a third scan the day before or after. ISCD recommended techniques were used to calculate same-day and consecutive-day precision error and LSC values. Results: There was high association between whole body (R2=0.98-1.00) and regional measures (R2=0.95-0.99) for same-day (R2=0.98-1.00) and consecutive-day (R2=0.95-0.98) measurements. The consecutive-day precision error, in comparison to same-day precision error, was significantly different (p<0.05), and almost twice as large for FM (1261g vs 660g), and over three times as large for LM (2083g vs 617g), yet still remained within the ISCD minimum acceptable limits for DXA precision error. No whole body differences in precision error were observed based on gender. Conclusion: When tracking changes in body composition, the use of precision error and LSC values calculated from consecutive-day analysis is advocated, given this takes into account both technical error and biological variation, thus providing a more accurate indication of true and meaningful change.

Details

Metrics

63 File views/ downloads
323 Record Views

InCites Highlights

These are selected metrics from InCites Benchmarking & Analytics tool, related to this output

Collaboration types
Domestic collaboration
International collaboration
Web Of Science research areas
Endocrinology & Metabolism

UN Sustainable Development Goals (SDGs)

This output has contributed to the advancement of the following goals:

#2 Zero Hunger
#3 Good Health and Well-Being
#5 Gender Equality

Source: InCites

Logo image