Journal article
Precision of the GE Lunar total body-less head scan for the measurement of three-compartment body composition in athletes
Journal of Clinical Densitometry, Vol.25(4), pp.692-698
2022
PMID: 36137876
Abstract
Introduction:
Dual energy X-ray absorptiometry (DXA) is widely used for the assessment of lean mass (LM), fat mass (FM) and bone mineral content (BMC). When observing standardised protocols, DXA has a high level of precision for the assessment of total body composition, including the head region. However, including the head region may have limited relevance in athletes and can be problematic when positioning taller athletes who exceed scan boundaries. This study investigated the precision of a new total-body-less-head (TBLH) DXA scan for three-compartment body composition measurement in athletes, with outcomes compared to the standard total-body DXA scan.
Methods:
Precision errors were calculated from two consecutive scans with re-positioning (Lunar iDXA, GE Healthcare, Madison, WI), in male and female athletes from a range of sports. TBLH precision was determined from repeat scans in 95 athletes (male n = 55; female n = 40; age: 26.0 ± 8.5 y; body mass: 81.2 ± 20.5 kg; stature: 1.77 ± 0.11 m), and standard total-body scan precision was derived from a sub-sample of 58 athletes (male n = 19; female n = 39; age: 27.6 ± 9.9 y; body mass: 69.6 ± 14.8 kg; stature: 1.72 ± 0.94 m). Data from the sub-sample were also used to compare precision error and 3-compartment body composition outcomes between the standard total-body scan and the TBLH scan.
Results:
TBLH precision errors [root mean squared-standard deviation, RMS-SD (coefficient of variation, %CV)] were bone mineral content (BMC): 15.6 g (0.5%), lean mass (LM): 254.3 g (0.4%) and fat mass (FM): 199.4 g (1.3%). These outcomes compared favourably to the precision errors derived from the standard total-body scan [BMC: 12.4 g (0.4%), LM: 202.2 g (0.4%), and FM: 160.8 g (1.1%)]. The TBLH scan resulted in lower BMC (-19.5%), LM (-6.6%), and FM (-4.5%) compared to the total-body scan (BMC: 2,308 vs. 2,865 g; LM: 46,954 vs. 50,276 g; FM: 15,183 vs. 15,888 g, all p<0.005).
Conclusion:
The TBLH scan demonstrates high in-vivo precision comparable to that of the standard total-body scan in a heterogeneous cohort of athletes. Given the impact of head exclusion on total body composition outcomes, TBLH scans should not be used interchangeably with the standard total-body scan.
Details
- Title
- Precision of the GE Lunar total body-less head scan for the measurement of three-compartment body composition in athletes
- Authors
- W Jones (Author) - Durham UniversityA Pearson (Author) - Durham UniversityD Glassbrook (Author) - Durham UniversityGary Slater (Author) - University of the Sunshine Coast, Queensland, School of Health and Behavioural Sciences - LegacyC Dodd-Reynolds (Author) - Durham UniversityK Hind (Corresponding Author) - Durham University
- Publication details
- Journal of Clinical Densitometry, Vol.25(4), pp.692-698
- Publisher
- Elsevier BV
- Date published
- 2022
- DOI
- 10.1016/j.jocd.2022.08.008
- ISSN
- 1559-0747
- PMID
- 36137876
- Copyright note
- Copyright 2022 The Authors. Published by Elsevier Inc. on behalf of The International Society for Clinical Densitometry. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/)
- Organisation Unit
- University of the Sunshine Coast, Queensland; School of Health - Nutrition & Dietetics; School of Health and Behavioural Sciences - Legacy
- Language
- English
- Record Identifier
- 99666998602621
- Output Type
- Journal article
Metrics
349 File views/ downloads
51 Record Views
InCites Highlights
These are selected metrics from InCites Benchmarking & Analytics tool, related to this output
- Collaboration types
- Domestic collaboration
- International collaboration
- Web Of Science research areas
- Endocrinology & Metabolism