Logo image
A low-cost thin layer coulometric microfluidic device based on an ion-selective membrane for calcium determination
Journal article   Peer reviewed

A low-cost thin layer coulometric microfluidic device based on an ion-selective membrane for calcium determination

Denis Dorokhin, G A Crespo, M G Afshar and Eric Bakker
The Analyst, Vol.139(1), pp.48-51
2014
url
https://doi.org/10.1039/C3AN01715JView
Published Version

Abstract

Analytical Chemistry Other Chemical Sciences thin layer coulometric microfluidic device ion-selective membrane calcium determination
A prototype of a low-cost and easy-to-use thin layer coulometric microfluidic device based on an ion-selective membrane for calcium detection is described. The microfluidic device was fabricated and consequently assembled with inexpensive materials without using sophisticated and centralized fabrication laboratory facilities. The linear range of the device is found to be 10-100 μM for a 60 s current integration time. Preliminary validations showed that the microfluidic device is suitable for the quantification of calcium in mineral water.